BetaCounter
Default conditions and parameters (WAS3ABi):
- Double-Side Silicon Strip detector (DSSSD): Micron BB7 (2mm pitch) / BB28 (1mm pitch)
- Size: 64 x 64 mm2
- Thickness: 1 mm (typical)
- Energy threshold: 20 ~ 30 keV (typical) & to be checked with the C2URIE)
- Dynamic range: 20 keV ~ 2 MeV for beta & 200 keV ~ 20 MeV for RI (adjustable)
- Number of channels: 256 channels or more
- Detection Efficiency: 60 - 80 % (typical) ... Depending on the implantation depth in DSSSD
- Beam loss from F11 to beta counter: 80-90% (typical), depending on beam focus and reactions in Al-degraders.
- RI implantation rate: 100 cps (typical) ~ 300 cps (maximum) in total
- The charged particle veto detector (plastic scintillation detectors): To be installed after the DSSSDs.
- Shield box: Al plate (0.2 mm thickness)
- Cooling: Optional
- Others: The DSSSDs are expected to be provided by users.